TECHNICAL PARAMETERS
Dimension17"X20"X 14"
Weight35 KGS
Voltage220-240V,50-60Hz,2A
Power140W
LeakageCurrent<0.1mA,220V
AmbientTemperature5-40oC
RelativeHumidity10 -80%
DISPLAY6" LCD Touch Screen with Blue Back Light. Complete function on LCD Display
PRINTERBuilt in Dot Matrix Printer
WATER REQUIREMENT
PurityR.O. Water in accordance with latest AAMI/ ANSI Standard for haemodialysis
Inlet Pressure15 psi - 60 psi
Inlet Temperature10 - 40°C
InletFlow2Ltr/min
WaterConsumption8 Ltr 1cycle
DrainageBelow Base of Machine, unobstructed, open to atmosphere
DISINFECTANT
Renaclean Or Any
Disinfectant Of Choice
Having Peracetic Acid
Concentration3.5 - 4.5%
Volume Required26 ml PER Cycle
Max. Pump Distance1 METER
DIALYZER TESTING
Pressure Test300mmHg (:1:10mmHg) pressure applied on the blood side for 30 Sec. Pressure drop should be less than 0.80mmHg/Sec. for low flux dialyzer and 1.20mmHg/Sec. for high flux dialyzer to pass the pressure test:
Volume Fail LimitAutomatically set at 80% of dialyzer's original priming volume.
Volume Test Scope0-150 ml (accuracy:l: 3ml)
REPROCESSING TIME
12 mins for one dialyzer
21 mins for two dialyzers at a time.