6" LCD Touch Screen with Blue Back Light. Complete function on LCD Display
PRINTER
Built in Dot Matrix Printer
WATER REQUIREMENT
Purity
R.O. Water in accordance with latest AAMI/ ANSI Standard for haemodialysis
Inlet Pressure
15 psi - 60 psi
Inlet Temperature
10 - 40°C
InletFlow
2Ltr/min
WaterConsumption
8 Ltr 1cycle
Drainage
Below Base of Machine, unobstructed, open to atmosphere
DISINFECTANT
Renaclean Or Any
Disinfectant Of Choice
Having Peracetic Acid
Concentration
3.5 - 4.5%
Volume Required
26 ml PER Cycle
Max. Pump Distance
1 METER
DIALYZER TESTING
Pressure Test
300mmHg (:1:10mmHg) pressure applied on the blood side for 30 Sec. Pressure drop should be less than 0.80mmHg/Sec. for low flux dialyzer and 1.20mmHg/Sec. for high flux dialyzer to pass the pressure test:
Volume Fail Limit
Automatically set at 80% of dialyzer's original priming volume.